基于身体质量指数优化管电流降低256层螺旋CT的冠状动脉成像辐射剂量的研究()

《医学研究与战创伤救治》(原医学研究生学报)[ISSN:1672-271X/CN:32-1713/R]

卷:
第16卷
期数:
2014年01期
页码:
1-5,9
栏目:
出版日期:
2014-02-20

文章信息/Info

Title:
The study of optimization of tube current based on BMI to reduce 256-slice CT coronary angiography radiation dose
作者:
雷 勇1陈自谦2付丽媛3钱根年3
1.350025 福建福州,南京军区福州总医院医务部;2.350025 福建福州,第二军医大学福州总院临床医学院医学影像中心;3.350025 福建福州,南京军区福州总医院医学影像中心
Author(s):
LEI Yong1CHEN Zi-qian2FU Li-yuan3QIAN Gen-nian3
1.Department of Medical Affairs,Fuzhou General Hospital of Nanjing Military Command,Fuzhou,Fujian 350025,China; 2.Medical Image Center,Clinical Medical College of Fuzhou General Hospital,The Second Military Medical University,Fuzhou,Fujian 350025,China;3.Medical Image Center,Fuzhou General Hospital of Nanjing Military Command,Fuzhou,Fujian 350025,China
关键词:
256层螺旋CT冠状动脉成像回顾性心电门控技术身体质量指数优化管电流图像质量辐射剂量
Keywords:
256-slice spiral computed tomography coronary angiography retrospective ECG-gating technique BMI optimization of tube current image quality radiation dose
分类号:
R816.2;R541.4
DOI:
-
文献标志码:
A
摘要:
目的 探讨基于身体质量指数(BMI)优化管电流在256层螺旋极速CT冠状动脉检查降低辐射剂量中的可行性与应用价值,即常规回顾性心电门控技术与基于BMI优化管电流回顾性心电门控术的对比研究。方法 按拟定的标准将120例病例纳入本研究,随机分为A、B两组,每组60例。A组:常规回顾性心电门控组(对照组);B组:基于BMI优化管电流回顾性心电门控组(实验组)。记录每例患者的性别、年龄、体重、身高、BMI、平均心率及扫描长度,测量并计算出原始轴位冠状动脉增强图像的图像噪声、信号噪声比和对比噪声比,同时进行图像质量评价,并计算出有效辐射剂量。然后对客观与主观图像质量评估指标及有效辐射剂量进行统计学分析。结果 ①A、B两组患者的图像噪声分别为:30.48±5.60、32.08±3.56,组间比较无显著差异(P>0.05)。A、B两组患者的图像信噪比分别为:15.42±3.73、14.47±2.72,组间比较无显著差异(P>0.05)。A、B两组患者的图像对比噪声比分别为:13.29±3.52、12.40±2.60,组间比较无显著差异(P>0.05)。A、B两组患者的图像总评分分别为:(3.63±0.41)分、(3.58±0.44)分,组间比较无显著差异(P>0.05)。②A、两组患者的有效辐射剂量分别为:(14.34±0.92)mSv、(7.16±1.40)mSv,组间比较有显著差异(P<0.05)。结论 基于BMI优化管电流回顾性心电门控与常规回顾性心电门控技术相比,前者在满足诊断要求的图像质量的同时,可有效地降低辐射剂量。
Abstract:
Objective To investigate the feasibility and application value of optimization of tube current based on BMI to reduce 256-slice CT coronary angiography radiation dose.Methods According to the exclusion criteria,120 cases were included in the study and divided into two groups randomly.Group A:conventional retrospective ECG-gating group (control group),group B:BMI-based optimization of tube current retrospective ECG-gating group (experiment group).Record the gender,age,weight,height,BMI,average heart rate and scanning length of each patient,and measurement background noise (BN),signal noise ratio (SNR) and contrast noise ratio (CNR) in axial original coronary artery enhancement image,assessment of the image quality of coronary artery subjectively,and calculated effective dose (ED).Then the basic information,the objective and subjective image quality assessment index and ED of the patients in the two groups were analyzed statistically.Results ① The BN of image in A and B groups were 30.48±5.60 and 32.08±3.56,no significant differences between the two groups (P>0.05).The SNR of image in A and B groups were 15.42±3.73 and 14.47±2.72,no significant differences between the two groups (P>0.05).The CNR of image in A and B groups were 13.29±3.52 and 12.40±2.60,no significant differences between the two groups (P>0.05).The total score of image in A and B groups were 3.63±0.41 and 3.58±0.44,no significant differences between the two groups (P>0.05).②The ED of patients in A and B groups were (14.34±0.92)mSv and (7.16±1.40)mSv,significant differences between the two groups (P<0.05).Conclusion Compared conventional retrospective ECG-gating CT coronary angiography with BMI-based optimization of tube current retrospective ECG-gating CT coronary angiography,the BMI-based optimization of tube current retrospective ECG-gating CT coronary angiography can reduce radiation dose effectively in the case of holding quality of diagnostic image.

参考文献/References:

[1] 王一民,曹建新,杨 诚,等.低管电压对双源CT冠状动脉成像图像质量和辐射剂量的影响[J].放射学实践,2010,25(9):999-1002.
[2] Leipsic J,LaBounty TM,Mancini GB,et al.A prospective randomized controlled trial to assess the diagnostic performance of reduced tube voltage for coronary CT angiography[J].Am J Roentgenol,2011,196(4):801-806.
[3] Ketelsen D,Fenchel M,Buchgeister M,et al.Estimation of radiation exposure of different dose saving techniques in 128-slice computed tomography coronary angiography[J].Eur J Radiol,2012,81(2):153-157.
[4] 李芳云,唐秉航,张晓东,等.256层CT前瞻性及回顾性心电门控ECG电流调制技术的应用[J].放射学实践,2011,26(7):781-784.
[5] Hou Y,Yue Y,Guo W,et al.Prospectively versus retrospectively ECG-gated 256-slice coronary CT angiography:image quality and radiation dose over expanded heart rates[J].Int J Cardiovasc Imaging,2012,28(1):153-162.
[6] Achenbach S,Marwan M,Ropers D,et al.Coronary computed tom-ography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition[J].Eur Heart J,2010,31(3):340-346.
[7] Hara AK,Paden RG,Silva AC,et al.Iterative reconstruction technique for reducing body radiation dose at CT:feasibility study[J].AJR,2009,193(3):764-771.
[8] Hirai N,Horiguchi J,Fujioka C,et al.Prospective versus retrospective ECG-gated 64-detector coronary CT angiography:assessment of image quality,stenosis,and radiation dose[J].Radiology,2008,248(2):424-430.
[9] Shuman WP,Branch KR,May JM,et al.Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries:comparison of image quality and patient radiation dose[J].Radiology,2008,248(2):431-437.
[10] Stolzmann P,Leschka S,Scheffel H,et al.Dual-source CT in step-and-shoot mode:noninvasive coronary angiography with low radiation dose[J].Radiology,2008,249(1):71-80.
[11] Ko SM,Kim NR,Kim DH,et al.Assessment of image quality and radiation dose in prospective ECG-triggered coronary CT angiography compared with retrospective ECG-gated coronary CT angiography[J].Int J Cardiovasc Imaging,2010,26(1):93-101.
[12] 张晓东,唐秉航,李芳云,等.心率对256层CT前瞻性心电门控冠状动脉成像质量的影响[J].放射学实践,2011,26(7):712-725.
[13] Austen WG,Edwards JE,Frye RL,et al.A reporting system on patients evaluated for coronary artery disease.Report of the Ad Hoc Committee for Grading of Coronary Artery Disease,Council on Cardiovascular Surgery,American Heart Association[J].Circulation,1975,51(4 Suppl):5-40.
[14] Bischoff B,Hein F,Meyer T,et al.Comparison of sequential and helical scanning for radiation dose and image quality:results of the prospective multicenter study on radiation dose estimates of cardiac CT angiography(PROTECTION) I Study[J].AJR,2010,194(6):1495-1499.
[15] Bongartz G,Golding SJ,Jurik AG,et al.European Commission Website.European guidelines for multislice computed tomography:appendix C.Funded by the European Commission,March 2004.Contract No.FIGM-CT2000-20078-CTTIP.www.msct.eu/PDF_FILES/EC%20CA%20Report%20D5%20% 20Dosimetry.pdf.Accessed March 26,2009.
[16] 吴兆苏,姚崇华,赵 冬,等.我国多省市心血管病趋势及决定因素的人群监测(中国MONICA方案)I.发病率和死亡率监测结果[J].中华心血管病杂志,1997,25(1):6-11.
[17] Brenner DJ,Hall EJ.Computed tomography:an increasing source of radiation exposure[J].N Engl J Med,2007,357(22):2277-2284.
[18] McCollough CH,Bruesewitz MR,Kofler JM Jr.CT dose reduction and dose management tools:overview of available options[J].Radiographics,2006,26(2):503-512.
[19] 高建华,孙宪昶,李剑颖,等.后置滤过器C2对64排螺旋CT冠状动脉成像质量的影响[J].中国医学影像技术,2007,23(1):73-76.

相似文献/References:

[1]梁永刚,付丽媛,陈自谦.综合评价法对256层CT前瞻性与回顾性心电门控冠状动脉成像质量的对比研究[J].医学研究与战创伤救治(原医学研究生学报),2015,17(06):563.[doi:10.3969/j.issn.1672-271X.2015.06.002]
 LIANG Yong-gang,FU Li-yuan,CHEN Zi-qian..Using comprehensive evaluation method to assess image quality of 256-slice CT coronary angiography in prospectively ECG-gating and retrospectively ECG-gating[J].JOURNAL OF MEDICALRESEARCH —COMBAT TRAUMA CARE,2015,17(01):563.[doi:10.3969/j.issn.1672-271X.2015.06.002]

备注/Memo

备注/Memo:
国家自然科学基金(81271607)
更新日期/Last Update: 2014-01-20