野战环境对Roche Cobas E411电化学发光免疫分析仪性能的影响()

《医学研究与战创伤救治》(原医学研究生学报)[ISSN:1672-271X/CN:32-1713/R]

卷:
第17卷
期数:
2015年06期
页码:
594-597
栏目:
出版日期:
2015-12-15

文章信息/Info

Title:
Influence of Roche Cobas E411performance under field environment
作者:
许永志陈 彬谢志雄郭健莲徐忠玉夏 挺
363000 福建漳州,解放军175医院检验科
Author(s):
XU Yong-zhi CHEN Bin XIE Zhi-xiong GUO Jian-lian XU Zhong-yu XIA Ting.
Department of Clinical Laboratory, 175 Hospital of PLA, Zhangzhou, Fujian 363000, China
关键词:
野战环境医学检验装备性能
Keywords:
field environment clinical laboratory equipment performance
分类号:
R446.6
DOI:
10.3969/j.issn.1672-271X.2015.06.010
文献标志码:
A
摘要:
目的 探讨野战环境对Roche CCobas E411电化学发光免疫分析仪性能的影响。方法 对Roche Cobas E411系统的降钙素原(PCT)、N末端脑钠肽前体(NT-proBNP)、肌红蛋白(Mb)、高敏肌钙蛋白T(hsTNT) 4个项目在野战环境下进行精密度、正确度、回收率、线性评价,并与Roche Cobas E601电化学发光分析仪测定进行比较。结果 该仪器测定PCT的低、高水平批内变异系数(CV)分别为1.70%、0.70%,批间CV为4.09%、3.35%;NT-proBNP的批内CV为2.46%、1.61%,批间CV为5.10%、2.59%;Mb的批内CV为2.42%、1.17%,批间CV为4.71%、2.51%;hsTNT的批内CV为2.83%、1.79%,批间CV为4.77%、3.14%。Roche Cobas E411与Roche Cobas E601电化学发光分析仪测定PCT、NT-proBNP、Mb、hsTNT的结果均显著相关(r>0.975, P>0.05),各项目回收率均在97%~103%。Roche Cobas E411测定PCT、NT-proBNP、Mb、hsTNT的线性范围分别为(0.08~98.75) ng/mL、(7.95~33815.65) pg/mL、(26.94~2816.40) ng/mL、(5.23~9311.05) ng/L。结论 Roche Cobas E411检测PCT、NT-proBNP、Mb、hsTNT具有优异的准确性、良好的精密度和较宽的线性范围,可满足野战环境需求。
Abstract:
Objective To explore the influence of Roche Cobas E411 electrochemistry luminescence immunity analyzer (Roche Cobas E411) performance under field environment. Methods Procalcitonin (PCT), N-terminal pro-B-type natriuretic peptide (NT-proBNP), myoglobin (Mb) and high sensitive cardiac troponin T (hsTNT) detected by Roche Cobas E411 system were evaluated in precision, accuracy, recovery and linearity under field environment. The data was compared with Roche Cobas E601. Results The within-run coefficient of variations (CVs) of PCT with low and high level control using Roche Cobas E 411 analyzer were 1.70%, 0.70% respectively, whereas the between-day CVs of PCT were 4.09%, 3.35%; the within-run CVs of NT-proBNP were 2.46%, 1.61%, which showed the between-day CVs of NT-proBNP were 5.10%, 2.59%; the within-run CVs of Mb were 2.42%, 1.17%, meanwhile the between-day CVs of Mb were 4.71%, 2.51%; and the within-run CVs of hsTNT were 2.83%, 1.79%, whereas the between-day CVs of hsTNT were 4.77%, 3.14%. The level of PCT, NT-proBNP, Mb and hsTNT with Roche Cobas E411 analyzer were well related to the results on Roche Cobas E601 (r>0.975, P>0.05), and the recovery of these four items were all between 97% and 103%. The linear range of PCT, NT-proBNP, Mb and hsTNT were (0.08-98.75) ng/mL, (7.95-33815.65) pg/mL, (26.94-2816.40) ng/mL and (5.23-9311.05) ng/L respectively. Conclusion Roche Cobas E411 analyzer presents good accuracy, outstanding precision and wide analytical measurement range, which can meet the requirement of field environment.

参考文献/References:

[1] 林 杰, 张建东, 刘高峰. 野战条件下医学检验的开展现状与对策[J]. 西南国防医药, 2012, 22(4): 440-441. [2] 杨永昌,王北宁. 野战医疗所检验装备及开展项目的现状和建议[J]. 人民军医, 2007, 50(6): 387. [3] 李卫东,陈礼潮.加强军队医院野战卫生装备训练的思考[J].东南国防医药,2014,16(2):221-222. [4] 王保军,周传华,龙 鹤,等.加强部队卫生装备使用管理的做法与体会[J].东南国防医药,2013,15(6):629-639. [5] Clinical and Laboratory Standards Institute.Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline-Second Edition[S]. EP5-A2, CLSI, 2004. [6] Clinical and Laboratory Standards Institute. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Second Edition[S]. EP9-A2, CLSI, 2002. [7] Clinical and Laboratory Standards Institute. Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Aproach; Approved Guideline[S].EP6-A, CLSI, 2003. [8] 傅占江. 野战检验医学装备现状及发展思路[J]. 医疗卫生装备, 2011, 32(2): 88-89. [9] 李 辉, 张祖进, 郭召平. 当前野战卫生装备存在的问题及改进意见[J]. 医疗卫生装备, 2011, 32(6): 77-78. [10] 包芙莉, 陈 强, 李 强, 等. 野战环境对检验医学及其装备的新要求[J]. 医疗卫生装备, 2011, 32(10): 116. [11] 林 杰,张建东,刘高峰.野战条件下医学检验的开展现状与对策[J].西南国防医药,2012,22(4):440-441. [12] 王 琦,胡爱民,王向荣,等.基于发展急诊医学提升军队医院机动卫勤救治能力的研究与思考[J].中国医药导报,2014,11(12):137-140. [13] 黄湘宁, 郑春苏. Roche Cobas e411全自动发光分析仪性能评价[J]. 检验医学与临床, 2011, 8(16): 2028-2029.